Measured response

In this view all measured points are plotted (and after evaluation also the smoothed curve shown in the Smoothed response view). The x axis is scaled in time [sec], and is usually logarithmic. The y axis is the voltage change on the measurement channel, converted to temperature difference [°C] using a calibration factor. This scale does not refer to absolute temperature, the lower end of the measurement range is shown as 0°C.


Figure 3-5: Measured curves in a multi-chip module, in still air chamber

Figure 3‑5 shows measured transients in a multi-chip module. Components are always measured in a certain measurement environment, in this case the module was mounted on a board of given size and placed in a closed chamber.
One of the chips had been powered by 200 mW for a prolonged time, the cooling was captured after switching off the power. Channel 0 shows the cooling of the chip powered, the other channels show heat transfer towards the other chips.

The number and distance of measured points is characteristic for the noise figure of the measurement, 2 points or 3 along a vertical line mean most measured values are the same, i.e. the noise is low.

Figure 3-6:  Detail view of Figure 3‑5

Figure 3‑6 demonstrates the quality of the measurement.